Uranyless Acetone

UranyLess in Acetone is a reagent used for cryo-electron microscopy applications, particularly for freeze substitution techniques (AFS). This product serves as a safer alternative to traditional uranyl acetate, providing a solution for biological sample preparation while ensuring high-resolution imaging.

Product Description:

  • Name: UranyLess Acetone
  • Applications: Freeze substitution (AFS), electron microscopy, cryo-applications. On block stain.
  • Format: 100 ml
  • Reference: 11000C-100
  • Usage Procedure: more information protocol below.
    1. Prepare the sample in an OsO₄ solution at -90°C for 10.5 hours.
    2. Increase the temperature for 0.5 hours.
    3. Immerse in an OsO₄ solution at -60°C for 10.5 hours.
    4. Wash with pure acetone at -30°C, then add UranyLess at -30°C for 8 hours.
    5. Gradually increase the temperature to 0°C to complete the protocol​.

This product is designed to maximize image resolution while minimizing risks to the user, offering an alternative to uranium-based reagents for specimen preparation.

Protocol AFS : UranyLess Acetone version cryo for AFS application stain sample on bloc

  • 8 hours 2% OsO4 solution at -90 °C
  • 0.5-hour temperature increase
  • 8 hours 2% OsO4 solution at -60 °C
  • 0.5-hour temperature increase
  • 8 hours 2% OsO4 solution at -30 °C
  • Remove OsO4 solution and add pure acetone
  • 1 hour washing
  • Remove acetone and add UranyLess acetone version cryo for AFS application.
  • 8 hours UranyLess acetone version at -30 °C
  • 1.5 hours temperature increase
  • 0 °C End of AFS

As soon as 0 °C was reached, the samples were washed three times with acetone and subsequently embedded in synthetic resin at room temperature.

plan for use uranyless cryo

With Uranyless Cryo:
membranes are clearly visible

With uranyless Cryo

Without Uranyless Cryo :
we didn't distinguish membranes

uranyless cryo result 2

Compare them

AqueousAcetone Ethanol 
Volume 30ml 200ml 100ml 30ml 200ml 
Reference 11000 11000-200 11000C-100 11000E-30 11000E-200 
Expiration12 Months 12 Months12 Months 12 Months 12 Months 
Use Contrast agent for ultrathin sections or negative stainingBlock staining for AFSStaining of dense and lipid-rich tissues
Cryofixation 
Technique Manual StainingAutomatic staining for EM stain Leica or RMC grids stainerManual StainingManual StainingAutomatic staining for EM stain Leica or RMC grids stainer
ContainerAirless bottle, no contact with airHDPE bottleGlass bottleAirless bottle, no contact with airHDPE bottle